application area
Widely used in fermentation, pharmaceuticals, food and brewingDomestic sewage treatmentSuitable for online process analysis and testing of small equipment cabinets, such as cooling water, drinking water, and purified water.
TECHNICAL INDEX
PH range |
0.00~14.00pH |
Power requirements |
16~36VDC |
PH resolution and accuracy |
0.01pH and ± 0.01pH |
power |
Max 4W |
Signal output load |
Max 600Ω |
Temperature compensation range |
0.0~100℃ |
ORP value range |
-1500~+1500mV |
Connection |
Pressing type terminal block |
PH mA signal output |
pH/ORP 4~20mA |
Shell material |
cast aluminium |
Temperature range |
-10~150℃ |
T-MA signal output |
Temperature range of 4-20mA |
Instrument size |
100*100*62mm |
IP GRADE |
IP67 |
ORP value resolution and accuracy |
1mV and ± 1mV |
ambient temperature |
-10~50℃ |
Instrument installation dimensions |
84*63mm |
Resolution and accuracy of temperature values |
0.1 ℃ and ± 0.3 ℃ |
storage temperature |
-20~70℃ |
Installation method |
Panel installation |
PH/ORP input |
Input impedance ≥ 10 Ω |
PT100/PT1000 2 or 3-wire system |
Display mode |
PH360 no display |
|
||
PH361 is displayed |
|
||||
|
|