YTS-60 desktop pneumatic pressure pump
The YTS-60 lever drives the large capacity chamber piston to make reciprocating motion, generating pressure or vacuum. The pressure or vacuum is outpu
Product details
2. Technical indicators
2.1 Pressure range: YTS-60: -0.095~6MPa
2.2 Sensitivity adjustment: min 10 Pa
2.3 Dimensions: 400 × 300 × 180 (length × width × height)
2.4 Weight: YTS-60: 8kg
3. Usage operation
3.1. Positive pressure detection
Press the toggle switch in, unscrew the plugs on the output 1, output 2, and output 3 connectors, connect the standard meter and the inspected meter, open the regulator valve, adjust the fine adjustment to the appropriate position (middle position), and close the air delivery valve. When the operating lever is pressurized to 4-5 MPa, use fine-tuning to adjust the pressure to around 6.5 MPa, stabilize for 1-2 minutes, use fine-tuning to adjust to the calibration point, close the stabilizing valve, and check point by point until zero.
3.2. Negative pressure detection
Pull out the toggle switch, unscrew the plugs on the output 1 and output 2 connectors, connect the standard meter and the inspected meter, open the stabilizing valve, adjust the fine adjustment to the appropriate position (middle position), and close the air delivery valve. Operate the lever to evacuate when the pressure reaches around -98kPa. Stabilize for 1-2 minutes, use fine-tuning to adjust to the calibration point, close the voltage regulator valve, and check point by point until zero.
3.3. When the pressure drops slowly under high pressure, close the stabilizing valve. (First, adjust the pressure to around 6.5MPa, stabilize it for 1-2 minutes, and then slowly adjust it to 6.0MPa)
4. Precautions
4.1 The calibration table should be placed on a flat and easy to operate surface, avoiding use in environments with highly corrosive gases and dust particles.
4.2 The tested instrument should be properly cleaned to prevent stains from entering the detection circuit.
4.3 Output 1 and Output 2 connectors are specialized quick couplings with sealed end faces,
When installing the instrument, there is no need to use any tools. Tighten it evenly, and when the sealing surfaces are tightly attached to each other, tighten it 1-2 turns.
4.4 The output 1 and output 2 connectors are composed of upper and lower parts. When using, it is necessary to regularly check whether the lower locking screw is connected reliably. If the connection is not reliable, it may cause breakage under high pressure.
When the pressure generated by the lever is heavy, please use a fine adjustment to increase the pressure, which will be easier.
5. Names and functions of each component

number name Function
1 Pump body assembly Combination installation of various components
2 Output 1 Connect standard table or inspected table
3 Output 2 Connect standard table or inspected table
4 Output 3 Connect standard table or inspected table
5 Piston Upper and lower piston device for generating gas
6 Air release valve Close when pressurized, release gas at the end of measurement
7 Leverage Drive the piston rod to move through operation
8 Switching valve Press in to measure positive pressure, pull out to measure negative pressure
9 Pressure regulator valve Lock when output pressure is unstable, can maintain stable pressure
10 Fine tuning Increase or decrease pressure within a small range