VDSeries/Universal Vacuum Oven
Dry safely and smoothly under vacuum. APT.line® The vacuum oven has unique performance and features, far ahead of similar products.
• Performance and features
-Electronic controlled APT.line®Internal preheating technology and two extended shelves
-Temperature range: from+15 ° C above ambient temperature to 200 ° C
-MP controller, with 2 programs (10 sections each) or 1 program (20 sections)
-The interval time between each section of a single program can be adjusted to a maximum of 99:55 hours or 999:59 hours
-The slope function can be adjusted through a program editor
-Passing time indicator
-Precise quantitative ventilation valve
-Anti shatter safety glass panel with springs installed
-Adopting cross flow technology and precise inert gas quantitative supply valve
-Independent adjustable temperature safety device, level 2 (DIN12880), equipped with a visual temperature alarm
-The rear has a direct 16mm measurement interface
-Simulated pressure gauge (displaying the pressure difference between the internal cavity and the surrounding environment)
-Inner cavity of electrolytic polishing
-All suction and ventilation pipes, as well as pressure vessels welded internally, detachable shelves, and vacuum valves, are made of stainless steel material 1.4571 (V4A)/AISI316 Ti
-Switching output of 2 × 24V DC (maximum 0.4 A), switched through 2 contacts in the program controller
-Used for switching software APT.COM®The RS422 interface of the data control system can be converted to the printer output port using an RS232/RS422 interface converter.
Technical specifications of VD series
VD23 |
VD53 |
VD115 |
|
Dimensions |
|||
Width (millimeters) |
515 |
635 |
740 |
Height (millimeters) |
665 |
775 |
900 |
Depth (millimeters) |
500 |
550 |
670 |
Observation window width (mm) |
250 |
350 |
456 |
Observation window height (mm) |
207 |
307 |
413 |
Distance between the rear and the wall (in millimeters) |
100 |
100 |
100 |
Distance between side and wall (mm) |
135 |
135 |
135 |
Internal dimensions |
|||
Width (millimeters) |
285 |
400 |
506 |
Height (millimeters) |
285 |
400 |
506 |
Depth (millimeters) |
295 |
340 |
460 |
Internal volume (liters) |
23 |
53 |
115 |
Aluminum extension shelves (standard quantity/maximum quantity) |
2/4 |
2/5 |
2/6 |
Load per shelf (kg) |
20 |
20 |
20 |
Allowable total load (kg) |
35 |
45 |
65 |
Weight (empty container) (kg) |
63 |
95 |
153 |
temperature data |
|||
Temperature range: 15 ° C to (° C) above ambient temperature |
200 |
200 |
200 |
Temperature deviation of 100 ° C(±°C) |
1.5 |
2 |
3.5 |
Temperature fluctuation range(±k) |
0.1 |
0.1 |
0.1 |
Heating time to 100 ° C (minutes) |
65 |
80 |
95 |
Vacuum interface with small flange (direct millimeter) |
16 |
16 |
16 |
Test hole with small flange (direct millimeter) |
16 |
16 |
16 |
Inert gas interface with flow restrictor (RP '') |
3/8 |
3/8 |
3/8 |
Allowable maximum vacuum degree (millibars) |
0.01 |
0.01 |
0.01 |
Vacuum leakage rate (maximum millibars/hour) |
0.01 |
0.01 |
0.01 |
Electric data |
|||
Protection level of enclosure in accordance with EN60529 standard |
IP20 |
IP20 |
IP20 |
Nominal voltage(±10%) 50/60 Hz (V) |
230 |
230 |
230 |
Nominal power (watts) |
800 |
1200 |
1700 |
Equipment safety device (A) |
10 |
10 |
10 |
All technologies are only applicable to 22±3 ° C ambient temperature and±Standard specification unloaded equipment under 10% power supply voltage fluctuation. Determine temperature data according to BINDER factory standards and DIN12880:2007, and adapt to the recommended wall gap of 10% of the internal chamber height, width, and depth. For the series, all data are typical averages. We reserve the right to make technical changes.