Beijing Jinkun Technology Co., Ltd
Home>Products>Ultra high sensitivity miniature spectrometer
Ultra high sensitivity miniature spectrometer
This spectrometer is a uniquely designed, single photon response level ultra high sensitivity miniature spectrometer. Due to its polarized optical des
Product details
Ultra high sensitivity miniature spectrometerFeatures: This spectrometer is uniquely designed with a single photon response levelUltra high sensitivity miniature spectrometerDue to its polarized optical design, unlike traditional grating spectrometers, it does not require slits or fibers, making it have a large light receiving area. Therefore, thisUltra high sensitivity spectrometerIt is an ideal choice for low light measurement and can be used for fluorescence spectroscopy, diffuse reflectance spectroscopy,Raman spectroscopy analysis.
ARCspectro HT is ideal for low light applications or for the spectral analysis of diffusing light sources (or samples). Its unique design permits togather a maximum of lightover a large diffusing surface (several millimeters!) and over a large angle (±10°). This gives atremendous throughput advantage(~50x more) compared to the most powerful grating spectrometers(also well-known Jaquinot advantage) which are intrinsically limited by a small entrance slit (typically 0.025mm).This makes this unique device ideal for applications were small light intensities has to be analyzed (fluorescence or Raman spectroscopy).

Ultra high sensitivity miniature spectrometerThe indicator parameters are shown in the following table:

Features ARCspectro HT-HR
Spectral range 400 nm – 1050 nm (HT-HR and HT-2D)

900-1700nm (HT-NIR)

Signal-to-Noise (single measurementfull detector dynamic used, Ratio between the peak value and bottom value of HeNe Spectrum) 300:1(Multiplex advantage)
Spectral Resolution <1 nm @l = 550 nm (HT-HR model)

<10 nm @l = 1000 nm (HT-NIR model)

Detector type SiliciumCCD front illuminateddetector array

InGaAs 512 pixels (HT-NIR) 1 stage cooled

Full-well detector capacity 30'000 e-
Dark noise 0.1% of dynamic range
Detector sensitivity ~15 photons per count
Relative wavelength fluctuation (measured with HeNe over 1 hour) < 0.01 nm
AbsoluteWavelength Accuracy < 0.5 nm
Operating Temperature 10°C – 50°C
Wavelength Drift with temperature 0.001nm per °C
Measurement time 1 ms to "infinite"
Operating Voltage USB powered
Light guide core diameter (optional) 4 mm
Entrance aperture 4 mm (Diam.)
Communication Interface USB2.0
Software Interface Windows XP, Window7, Labview
CCD Readout Sampling rate 0.5 MHz
A/D converter 16bits (8-10 bits for HT-2D)
Product Dimensions (including electronics) 15cm x 10cm x10cm
Product Weight 300 g