Shandong Lunan Ruihong Chemical Instrument Co., Ltd
Home>Products>SP-6802 Trace Sulfur Analyzer
Firm Information
  • Transaction Level
    VIP member
  • Contact
  • Phone
    18563203209
  • Address
    No. 1866 Hengyuan South Road, Tengzhou Economic Development Zone, Shandong Province
Contact Now
SP-6802 Trace Sulfur Analyzer
(1) Minimum pick-up limit (based on S in H2S): Type A machine (non conversion) 0.01 ㎎ s/m3; B-type machine (non conversion) 0.1 ㎎ s/m3; C-type machine
Product details

(1) Minimum Pick Up Limit (in H)2S in S): Type A machine (not converted) 0.01㎎s/m3 BType machine (non conversion) 0.1 ㎎ s/m3 C-type machine (converted) 0.2 ㎎ s/m3.

(2) Non dilution measurement range: 0.01-50 (0.1-500) μ s/m3(with H)2S in S calculation).

(3) Data repeatability: bottledH2SCS equipped with standard gas or instruments2The root mean square error of the analysis results is less than 5% after continuous injection analysis of the permeation source gas for more than 10 times.

(4) Column temperature: adjustable from room temperature to 100 ℃± 0.2 ℃.

(5) Permeation source temperature: adjustable from room temperature to 50 ℃± 0.1 ℃.

(6) Appraiser temperature: adjustable from room temperature to 110 ℃± 1 ℃.

(7) Converter temperature: adjustable from room temperature to 900 ℃± 5 ℃.

(8) High voltage power supply: - Adjustable from 400V to -800V.


Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!

Successful operation!