SF6Infrared dual wave quantitative leak detector(IAC-510)
FranceIACImported with original packaging, lightweight and precise;Powered by batteries, precise leak detection can be achieved freely in any situation! 
Overview
SF6Infrared dual wave quantitative leak detector(IAC-510)It uses the principle of infrared spectroscopy for detectionSF6Gas. The infrared wavelength emitted by infrared spectroscopy is only applicable toSF6Gases themselves have reactions and do not react with any other gases. Spectral absorption method is a method of detecting gas concentration by detecting changes in the transmitted or reflected light intensity of a gas. Each gas molecule has its own absorption(Or radiation)Spectral characteristics, the emission spectrum of the light source only produces absorption in the overlapping part with the gas absorption spectrum, and the absorbed light intensity will change, thus having a high degree of selectivity. So this instrument can also be used normally under anaerobic conditions. However, instruments based on the discharge principle of electrochemistry cannot.


Features

SF6The infrared dual wave quantitative leak detector is easy to use, portable, and can detect the smallest particles in the airSF6The main structural characteristics of gas concentration are its sensitivity, response time, and stability against humidity and toxic gases. Based on the principle of dual wave non dispersive infrared radiation, it provides reliable and high-precision performance with minimal maintenance.SF6The gas leak detector has significantly developed to be able to detect indoor and outdoor fillingSF6The small leakage of gas equipment is to reduce environmental pollution andSF6Ideal tool for gas leak detection.
l Powerful data storage capabilities.
l Data browsing function.
l RS232Easy to connect interfacePCPerform computer processing.
l Display: Timemg/m3ThePPNThree functional displays.
l Touch screen operation:300000time.


l Data browsing function.
l RS232Easy to connect interfacePCPerform computer processing.
l Display: Timemg/m3ThePPNThree functional displays.
l Touch screen operation:300000time.
l Unrestricted by humidity content.
l Particle filter for sensor protection.
l Powerful data storage capabilities.
l Data browsing function.
l RS232Easy to connect interfacePCPerform computer processing.
l Display: Timemg/m3ThePPNThree functional displays.
l Touch screen operation:300000Next time.
l Unrestricted by humidity content.
l Particle filter for sensor protection.
Technical Specifications
l Technical features: Infrared detection technology(NDIR)Double wave
l Gas detection:SF6
l stable fixed Sex:TLVMonitoring, high stability
l Service life: greater than10year
l Detection range:0-1000ppm
l essence Degree: Positive or negative of full range1%
l Operating Temperature-30Degree to50degree
l pressure range:800hPa-1000hPa
l Humidity range:0%-95%
l Preheating time: less than2minute
l Reaction time: diffusion time less than25Seconds, flow rate less than5second
l Storage temperature:-30to50degree
l Storage data:100article
l Host size:456×372×185mm
l Host weight:3Kg
l Handle size:200×60×180mm
l Handle weight:0.5Kg
Service Guarantee

l Provide on-site training guidance and conduct on-site demonstration operations.
l Provide detailed Chinese manuals and user guidance.
l One year free warranty, lifetime spare parts support and maintenance.
l hotline365Answer user inquiries 24/7.