PTV2000 online turbidity measuring instrument
Intelligent interface
Product details
The PTV200 new generation online turbidity measuring instrument determines turbidity as a key parameter for evaluating the quality of drinking water.
Intelligent interface
Simplify and optimize processes
Innovative design with low maintenance requirements
For the measurement of low pollution wastewater, it has high accuracy
Minimalist Operations - Innovative Methods
Everything comes from the emphasis on details: During the research and development process of the Roewe Bang PTV series products, full consideration was given to making all aspects of online turbidity monitoring simple and convenient, including simplifying installation, debugging, and setup; Daily operation and control; Conventional procedures such as calibration, verification, and maintenance, as well as data acquisition and management
Intelligent User Interface - Redefine Controller
Using familiar interfaces of smart devices such as smartphones and tablets for short-range wireless control, replacing traditional controllers. The mobile app has simple data retrieval and statistical analysis functions, quick access to operation guides and step-by-step prompts for help information, greatly improving and enhancing user experience, and simplifying operations
save resources
Using PTV instruments saves time, money, and water. The pursuit of low range and high precision in drinking water turbidity monitoring (especially in the case of<1NTU), PTV online turbidity meter is specially designed and optimized for such requirements, achieving unimaginable monitoring effects and helping users save money and time costs
Compliant with international standards and specifications
There are multiple versions of PTV online turbidity meters available for different countries and regions to choose from, covering ISO standards or US EPA standards
Inlet and outlet water management device
The inlet and outlet water management device is an additional device for managing water sample inlet and outlet PTV, integrating functions that require the use of multiple separate components, saving a lot of installation and maintenance work
Install the backboard
The installation backplate is an additional accessory used for quick and easy installation and fixation of major instrument components and related accessories
Two innovative ways to operate the interface
Touch screen operation interface - PTV series products come with a touch screen, which makes all settings and instrument operations very easy without the need for any external devices
AquaLXP for mobile devices ™ App operation interface - Use AquaLXP TM app operation software to operate PTV devices, which is an enhanced version of touch screen operation. Not only can you perform all routine setting operations on PTV devices, but you can also enjoy the intelligent heuristic animation operation guidance, advanced data analysis and statistical tools, complete maintenance and repair logs, and unparalleled comfort brought by familiar device use
Standard solution - used for calibration&validation
From Tintometer ® Lovibond ® Lovibond ® The T-CAL turbidity standard solution is a first-class standard solution that meets the requirements of US EPA and ISO standards T-CAL turbidity standard solution and T-CALplus ™ When used in conjunction with a calibration kit, you will find that turbidity calibration is so simple and safe.

Technical specifications of PTV2000 new generation online turbidity meter
Measurement parameters measuring range
Turbidity 0.0001 - 100 NTU
technical specifications
optical properties Red LED
Power supply requirements 90-264 V, AC, 50-60 Hz, automatic selection
accuracy ± 2% of the final range value or less than 0.008/8 mNTU
stray Less than 0008/8 mNTU
display resolution Up to 0.0001 NTU (depending on range) or 5 display bits
Repeatability/Accuracy At 1 NTU, it is better than 1%
initial response 10% change: 15 seconds at maximum traffic
Step response T-90), 200 milliliters per minute in less than 240 seconds. At 1NTU
signal averaging Optional: 1, 3, 6, 10, 30, 60, and 90 seconds, with a standard setting of 30 seconds
sample temperature 0-50 ° C, the highest sample temperature is 70 ° C
calibration Single point calibration using licensed formalin solution for 5.0 or 20NTU
sample flow 30 to 500 ml/min, optimal condition 50 to 80 ml/min
injection port 1/4 inch NPT female head, 1/4 inch compression hose (included)
Sample outlet (Drainage) 3/8-inch NPT female head, 3/8-inch hose connector (included)
Injection tube 1/4 inch OD or 6 mm OD
Sample tube 3/8 inch OD or 9 mm OD
Turbidity meter emptying Quick connector for integrated globe valve
Analog output: Measurement module You can choose between 0-20mA or 4-20mA; The output range can be programmed through each segment of the measuring range
Analog output: power box connection You can choose between 0-20mA or 4-20mA; The output range can be programmed through each segment of the measuring range
call the police Three set-point alarms, each equipped with an SPDT relay with unpowered contacts rated 5 A resistive load at 230 V, AC
Digital protocol 4-20 mA
Type of junction box casing Polyester fiber reinforcement
Protection level of junction box shell IP 66
safety Registered by T Ü V Rheinland, compliant with UL 61010A-1: certified by T Ü V Rheinland, compliant with CSAC22.2 No.1010.1: CE certified by T Ü V Rheinland, compliant with 61010-1
immunity CE certification by T Ü V Rheinland, compliant with 61326 (industrial standard)
Emissions Grade A: EN 61326, CISPR 11, FCC Part 15, Canadian standard for interference causing digital devices ICES-003
install hardware Turbidity meter sensor - slotted mounting bracket, can be fixed on any vertical plane or panel (optional). Junction box - directly installed on any vertical plane or panel (optional)
authentication method Wet standard (stable formalin solution) or dry detection device
Limit of Determination 0.0005 NTU
detection limit 0.001 NTU
environmental condition 5-50 ° C, relative humidity 5-95% (non condensing)
consistency EPA
protection grade IP 65
size 158.4 x 340.4 x 334.5 mm
weight 5 kg