Secondary Biosafety Cabinet BHC-1300IIA/B2 | Biosafety Cabinet Modelsummary:
Biological clean safety cabinets are essential items in today's microbiology laboratories, especially in situations where protective measures need to be taken for operators. A sterile, dust-free, and safe working environment for bacterial cultivation in medical, pharmaceutical, scientific research, and other fields.
Secondary Biosafety Cabinet BHC-1300IIA/B2 | Biosafety Cabinet ModelCharacteristics:
1. Air curtain isolation design to prevent internal and external cross contamination. 70% of the airflow is discharged externally and 30% is circulated internally, with negative pressure vertical flow.
2. The glass door can be moved up and down freely, easy to operate, and can be closed for sterilization. The positioning height limit alarm prompt.
3 power output sockets in the workspace, equipped with waterproof sockets and drainage interfaces for easy operation
Special filters are installed at the exhaust outlets to control emissions and pollution
The working environment is made of 304 stainless steel, which is smooth, seamless, and has no dead corners. It can be easily disinfected to prevent corrosion by corrosive agents and disinfectants.
6 adopts LED LCD panel control.
Equipped with high-efficiency filter failure alarm function.
8 Work area airflow leakage alarm.
Technical parameters:
model |
BHC-1300IIA/B2 |
Cleanliness |
Level 100 @ ≥ 0.5 μ m (Federal 209E) |
cfu |
≤ 0.5 pieces per dish Time (90mm diameter petri dish) |
noise |
≤62dB(A) |
Average wind speed inside the door |
0.38+-0.025m/s |
Average wind speed in the middle |
0.26+-0.025m/S |
Average wind speed on the inner side |
0.27+-0.025m/S |
Vibration half peak value |
≤ 0.5um (x, y, z direction) |
illumination |
≥300LX |
Front suction wind speed |
0.35+-0.025m/s |
Air tightness |
<=10-6m/S (under a pressure of 500Pa) |
Power Supply |
AC, Single phase 220V/50H |
High power consumption |
1000W |
weight |
220KG |
Working size |
1000*675*620mm |
External dimensions |
1195*735*1950mm |
Specification and quantity of high-efficiency filters |
995×640×38×1mm |
Specification and quantity of fluorescent/ultraviolet lamps |
20W*1/20W*1 |