SX-L3100T PharmaceuticalHigh flow laser dust particle counter
It can directly detect clean environments with cleanliness levels of * zone and B zone. The SX-L3100T laser dust particle counter adopts a semiconductor laser light source and a large touch screen display, providing simple operation mode, fast sampling, lightweight and easy to carry characteristics, effectively saving the operator's time. The lightweight and streamlined stainless steel casing, combined with ergonomic handles, allows users to easily lift and place it in any suitable measuring location with one hand. Additionally, data can be downloaded and stored through computer software.
The product is widely used for clean environment testing in industries such as pharmaceuticals, optics, chemistry, food, cosmetics, electronic health, biological products, aerospace, etc.
SX-L3100T PharmaceuticalHigh flow laser dust particle counter
Technical Specifications
| sampling flow | 100L /min(3.53cfm) |
| Particle size channel | 0.3,0.5,1,3,5,10μm |
| cleanliness recovery characteristic | ≤10min |
| Particle size distribution error | 0.5μm, Particle size distribution error ± 30% |
| Error in particle size concentration indication | After the particle counter is in normal working condition, the error of particle concentration indication in the 0.5 micron particle size range shall not exceed ± 10%. |
| Repetitive relative deviation | Under the same measurement conditions, the repeatability of continuous particle concentration measurements is ± 10% |
| Large sampling concentration | 35000 pieces/liter |
| Cleanliness level determination criteria | ISO14644-1, GMP dynamic, GMP static |
| laser light source | laser diode |
| Imported air pump | Continuous and stable operation, energy-saving and environmentally friendly |
| Data communication interface | USB |
| temperature and humidity | optional |
| power supply | AC 220V 50Hz |
| Built-in lithium battery | DC 14.8V 36Ah |
| Charging adapter | Input 100V~240V 2.5A, output 16.8V 5A |
| Battery charging time | 8 hours |
| External dimensions/weight | Length 291 x Width 265 x Height 280 (mm)3)/9.15kg |
| material | SUS304 stainless steel |
| Usage environment | 0~40℃,10-90%RH |
| Stored Environment | Usage environment: Temperature 0~40 ℃, relative humidity 10~70% PH Storage environment: Temperature -30~45 ℃, relative humidity 0~90% PH |
| working hours | Fully charged for continuous testing time of 4 hours |
| sampling time | User self setting (1s-14400s) |
| SAMPLE DELAY | User self setting (1s-255s) |
| sampling times | 2~50 times |
| number of sampling points | 2-9 times |
| Regional location | From 0 to 99 |
| Data memory capacity | 1~5000 sets of measurement data, searchable |
| count mode | cumulative number |
| screen display | Colorful 7-inch touch screen |
| Printing method | Built-in Thermal Printer |
| Verification standards | GB / T 6167 - 2007 / JJF 1190 - 2008 |
| Standard Appendix | Sampling port rubber sleeve, equal power sampling head, self-cleaning filter, tripod, printing paper, portable protective box, charging adapter, export data CD, USB data cable, anti-static black hose (inner diameter 10mm), user manual, original factory calibration report, certificate of conformity. |
| Audit tracking function | Compliant with FDA 21CFR PART 11 requirements, involving access security, audit trails, data storage and retrieval, and electronic signatures. |