The airborne microbial testing is an active sampling method that is not selective for microorganisms and can be quantified.MERCK MilliporecompanyMAS-100NT®andM air T®The series is designed based on the classic Anderson impact principle, where microorganisms in the air are "inoculated" by hitting the agar surface through a perforated cover (sieve). Remove the culture medium and culture it under suitable conditions to count the bacterial colonies. Two series of instruments fully comply withISO14698Validation standards, with efficient physical and biological efficiency, fast sampling speed, small air volume error, and high microbial recovery growth rate. Additionally, provide a completeIQ/OQVerify text and services with professional verification and repair services.
MAS-100NT® Item No.1.09191.0001
¨ use
¨ Accurate measurement, built-in anemometer for flow compensation
¨ SQSContinuous Fractional Sampling System
¨ 4Minute automatic verification system
¨ Display the time and date of each sampling, data storage andUSBTransmission function
¨ Adjustable sampling angle
¨ Wind speed alarm
Technical Specifications:
Material: anodized aluminum;
Height:25cm
Diameter:11cm
Weight:2.38kg
Sampling head diameter:10cm
Current Speed:100rise/Score, accuracy±2.5%, in accordance withIso14698-1standard
Airflow regulation: Built in wind speed flowmeter automatically adjusts flow rate
Sampling volume: Preset volume:50,100,250,500,1000Raise. Each volume can be preset as1-2000Raise.
Petri dish holder:90mmDiameter standard culture dish
Rechargeable battery: lithium battery pack,7.4V/6.9Ah
Power supply:110-240V,50-60KHz
Motor:6Vdirect current
Display screen: high-definition white background chart display screen
Battery life:Real time clock, lifespan10year
Operating conditions: Temperature0—40℃, humidity0—80%(relative humidity)
CECertificationEmission:EN 61326-1:2006, EN55011:1998+A1:99,
Immunity: EN 61326-1:2006,EN 6
EN 6
EN 6
EN 6
Data cable interface:USB 2.0