Adopting standards:
JJG956-2013 "Verification Regulations for Atmospheric Samplers"
JJG943-2011 "Total Suspended Particle Sampler"
HJ/T374-2007 Technical Requirements and Testing Methods for Total Suspended Particle Samplers
HJ/T375-2007 "Technical Requirements and Testing Methods for Environmental Air Samplers"
Q/2013LDZ 006-2015 "Atmospheric Comprehensive Sampler"
Technical features:
1. Can collect gas pollutants and TSP separately or simultaneously, with optional PM10/PM5/PM2.5 sampling probes;
2. Aluminum alloy particle sampling head effectively eliminates electrostatic adsorption;
3. Efficient and low-noise brushless fan;
4. OLED screen display, Chinese menu operation, wide temperature operation;
5. The atmospheric part adopts dual constant current sampling, with dual Japanese diaphragm pumps controlled separately;
6. The motor has intelligent automatic protection function. If the particulate matter sampling still does not reach the set flow rate within 35 seconds and the pressure is less than -9.0KPa, the sampling will automatically stop and wait for 30 minutes before starting again;
Automatic memory of 7 sampled data, providing 80 sets of data (expandable) for users to query and print; Equipped with a real-time clock, supporting timed and interval sampling settings;
8. Power down protection data, automatic recovery of sampling upon incoming call; Overload automatic protection function;
9. Directly measure atmospheric pressure and automatically participate in standard condition volume calculation;
10. One click sampling with user-friendly design, no need to reset parameters when restarting;
11. The heater is located inside the main unit and comes with an insulation cover. It has a light shielding function and can collect gas in low-temperature environments;
12. Equipped with a sampling gas path anti backflow device and capable of effectively achieving dehumidification protection function;
13. Password protection function ensures secure operation;
14. USB drive support.
Main indicators:
|
Main parameters |
parameter range |
resolution |
accuracy |
|
Sampling flow rate of particulate matter |
(60~125)L/min |
0.1 L/min |
Better than ± 2% |
|
Atmospheric sampling flow rate |
(0.1~1)L/min |
0.1 L/min |
Better than ± 2.5% |
|
sampling time |
1f minutes~99 hours and 59 minutes (adjustable) |
0.01L/min |
Better than ± 0.2% |
|
Pre measurement temperature |
(-30~99)℃ |
0.1℃ |
Better than ± 2 ℃ |
|
atmospheric pressure |
(70~130)kPa |
0.1kPa |
Better than ± 2.5% |
|
sampling times |
unrestricted |
||
|
Control method |
Automatic; Manual: on/off, unlimited sampling time |
||
|
Size |
(Length 200 x Width 255 x Height 320) mm |
||
|
Total weight |
About 7Kg |
||
|
working environment |
Temperature (-20~50) ℃; Humidity ≤ 95%; Atmospheric pressure (80-110) KPa |
||
|
working power supply |
AC220V±22V,50Hz |
||
|
Power consumption and noise |
<550W、≤62dB |
||