The 8400D explosion-proof online gas analyzer adopts an explosion-proof structural design and is suitable for hazardous areas. It can integrate various detection technologies such as infrared, electrochemical, and magnetic oxygen, and can measure up to 270 types of gases. We can customize corresponding instruments according to customer needs, and one instrument can measure up to 6 gases simultaneously. Not only can the types of gases measured be customized according to needs, but instrument configurations can also be customized according to customers' accuracy and cost requirements, with great flexibility and applicability.


Working Principle:
The IR-208 adopts a modular design and can integrate multiple detection technologies such as NDIR infrared and electrochemical to analyze various gas components.
Most gases can absorb infrared light of different wavelengths and have unique infrared spectral absorption characteristics. The light attenuation caused by absorption satisfies Lambert Beer's law, according to which the concentration of a gas is proportional to the intensity attenuation of specific wavelengths of infrared light.
When the infrared light generated by the infrared light source of the instrument passes through the measuring chamber, the measured gas in the chamber will absorb infrared light of a specific wavelength, causing it to attenuate. The intensity of the infrared light at that frequency is detected by a detector on the other side. By comparing the intensity of the reference detector, the attenuation coefficient of the infrared light at that wavelength can be determined. According to Lambert Beer's law, the gas concentration can be calculated.

Product Features
A single analyzer can measure up to 6 gases
High measurement accuracy, up to 1% FS
Built in temperature pressure flow detector, automatic temperature pressure compensation
Adopting stainless steel measuring cell and gold-plated reflective coating, sapphire light window
Corrosion resistant material, good adaptability, high stability
Large screen LCD display, non-contact keyboard operation
Over 50 years of experience in designing and manufacturing infrared analyzers, with stable and reliable quality
Customized design and manufacturing technology to provide customers with the most suitable products
Explosion proof structure, suitable for hazardous zones one and two
Low maintenance and long instrument lifespan
Application fields:
Oil and gas transportation and recycling
Continuous Emission Monitoring System (CEMS)
Explosion limit LEL monitoring
Industrial Process Analysis
technical indicators
Measurement indicators: THC, CH4 and other gases, can be customized according to needs
Testing methods: NDIR, electrochemistry, magnetic oxygen, etc
Measurement range: 0-100%, customizable range
Response time:<10 seconds
Repeatability: ± 1% FS or ± 2% Reading
Measurement accuracy: ± 1% FS or ± 2% Reading
Zero drift: ± 1% FS/90 days
Range drift: ± 2% FS/90 days
Linearity: ± 1% FS
Resolution: ≤ 1ppm
Output: Isolated 4-20mA, 0-1/5VDC (optional), RS232
Alarm output: High and low limit alarm 9 (optional)
Power supply: 90-260VAC, 50/60Hz,<3 Amp
12, 24, or 36VDC (optional)
Sample contact material: 316SS measuring cell, sapphire light window, PTFE
Sample flow rate: greater than 0.5L/min, less than 2L/min
Sample temperature: 0-70 ° C
Sample pressure:<5psi, automatic pressure compensation
Sample conditions: no condensation, no dust
Preheating time: 5 minutes available, 15 minutes completely stable
Environmental temperature: -40 to 50 ° C
Environmental humidity:<95% RH
Air interface: 1/4 "
Explosion proof rating: Exd IIB T4, IP66
Size: 10.25 "x 10.25" x 9 "(26 cm x 26 cm x 22.86cm)
Weight: 25 lbs. (11.35 kg)