The Herocell 240 CO2 incubator, based on years of production experience and design breakthroughs, has demonstrated some key characteristics in practical cell culture applications. Providing the optimal cell growth environment and effective pollution control technology, the thoughtful design is suitable for more important applications, and it is easy to operate and monitor, allowing you to invest more time in your research goals.
Herocell 240 CO2 incubatorAdopting a 6-sided direct heating temperature control method, it can provide precise temperature control for your cell culture environment, with a temperature uniformity of up to±0.1℃;140°The dry heat high-temperature sterilization function can easily achieve complete sterilization with one click, effectively avoiding the risk of bacterial contamination in cell culture;Infrared (IR)CO2Concentration detector accurately controls CO inside the box2Concentration; The touch control panel allows for easy viewing of historical data curves and is equipped with a USB interface for one click export of historical data. Our reliable direct heating carbon dioxide incubator adopts a unique HEPA air filtration system built-in to the cabin, which can sustainably protect the cultivation environment from harmful pollutants in the air. It provides high-temperature sterilization as needed to simplify daily cleaning and maintenance.
Product advantages:
❏6-sided heated direct heating compartment
▸The 248L large capacity compartment provides ample cultivation space and an ideal environment for cell culture applications
▸6-sided heating method,The efficient and high-performance heating system distributed on the surface of each cultivation chamber provides a highly uniform temperature distribution for the entire incubator,Make the temperature of the entire incubator more uniform, and stabilize the temperature field inside the chamber to achieve a high degree of uniformity±0.1℃
▸ Standard right side door opening method, optional left and right door opening direction according to requirements
▸ Polished stainless steel integrated inner cavity rounded corner design for easy cleaning
▸ Removable pallet flexible combination, independent humidity tray can be removed or placed according to needs
▸ The built-in fan in the cabin gently blows air to evenly distribute it inside the cabin, ensuring a consistent cultivation environment
▸ Stainless steel partitions and brackets are sturdy and durable, and can be disassembled without the need for tools
❏Non condensing water technology
▸Non condensing water control technologyUnder normal room temperature conditions, even when the humidity tray generates high humidity, condensation water will not be generated above the cabin, providing maximum protection for cell and tissue culture and avoiding the danger of condensation formation. Turbulent chamber ventilation ensures a constant and uniform cell culture environment
❏ 140°CDry heat high-temperature sterilization
▸ Provide as needed140°CDry heat high-temperature sterilization simplifies cleaning work, eliminating the need for separate high-temperature and high-pressure disinfection and reassembly of components, effectively improving work efficiency
The 140 ° C dry heat high-temperature sterilization system can effectively eliminate bacteria, mold, yeast, and mycoplasma on the surface of the inner cavity
❏ ISO Level 5 HEPA filtered airflow system
▸The built-in HEPA air filtration system in the cabin can continuously filter the air throughout the entire cabin
▸Within 5 minutes after opening the door, achieve ISO level 5 air quality
▸Provide continuous protection by reducing the adhesion ability of air pollutants on internal surfaces
❏ImportSensors and probes designed for precise monitoring
▸ Infrared (IR) CO2Sensors provide stable monitoring when humidity and temperature are not very predictable,Effectively avoids the measurement deviation problem caused by frequent opening and closing of doors
▸ Most suitable for sensitive applications and remote monitoring, or situations where frequent opening of the incubator is required
▸Temperature probe providing overheat protection
❏ProactivelyAirflow technology
▸The incubator is equipped with a fan assisted airflow circulation, which can achieve rapid recovery and strict consistency. Our airflow pattern is specifically designed to evenly distribute key environmental conditions such as temperature, gas exchange, and humidity
▸The built-in fan in the chamber can gently blow filtered moist air throughout the entire chamber, ensuring that all cells, regardless of their placement, have the same cultivation environment conditions and will not excessively lose water
❏ 5inchLCDTouch control screen
▸Intuitive control and easy operation,Can display real-time running curves and historical running curves
Located above the doorThe installation location is convenient and easy to control, and the capacitive touch screen is sensitive to touch and provides a good control experience
▸Sound and visual alarms,Screen menu prompt
❏ Historical data can be viewed, monitored, and exported
▸Historical data can be viewed, monitored, and exported through a USB interface. Historical data cannot be changed and can be effectively traced back to the original data
Product configuration:
model Herocell 240,The corresponding configuration is:
1 host, 1 HEPA filter, 1 air inlet filter, 1 stainless steel humidification tray, 3 stainless steel palletsCO2Supply intake pipe 1.5 meters1 power cord and 1 piece of paper material
Technical parameters:
typenumber | Herocell 240 |
Control interface | 5-inch LCD touch screen |
Temperature control mode | PID control mode |
Heating method | 6-sided straight hot air sleeve type |
Temperature control range | Room temperature+5 ℃~60℃ |
Temperature display resolution | 0.1℃ |
temperature stability | ±0.1℃ |
Uniformity of temperature field | ±0.3℃ (at 37℃) |
heating power | 900W |
Timer function | 0-999.9 hours |
Internal dimensions(Width x Depth x Height) | 630 x 600 x 670mm |
External dimensions(Width x Depth x Height) | 800 x 660 x 1000mm |
Box volume (L) | 248L |
CO2measuring principle | Infrared (IR) detection |
CO2Control range | 0-20% |
CO2Display resolution | 0.10% |
CO2supply |
Maximum 10bar (1MPa)Overvoltage Suggest using 0.5~1bar(0.05~0.1MPa) |
relative humidity (Humidification plate humidification) |
ambient humidity ~95% at 37℃ |
HEPA filtration | ISO Level 5, 5 minutes |
Sterilization method |
140°CDry heat high-temperature sterilization |
Temperature recovery time |
≤10 min
(Open the door for 30 seconds, room temperature 25)℃Set value 37℃)
|
Gas concentration recovery time |
≤5 min
(Open the door for 30 secondsSet value 5%)
|
Data export interface | USB interface |
Historical data storage | 25W pieces of information |
Working environment temperature | 10 ℃ to 30 ℃ |
Power Supply | 220~240V/50~60Hz |
weight | 116kg |
Scalability | Up to 2 units can be stacked |