Introduction to FD216 Environmental Radon Measuring Instrument:
FD216 Environmental Radon Measuring InstrumentIt is a measurement of radon concentration (optional) in ambient air, soil, water, and soil radon release rate (optional). It can be used for geological exploration, radiation protection, nuclear accident monitoring, radiation dose evaluation, earthquake prediction, and teaching.
This instrument complies with the measurement principles and requirements of GB50325-2010 "Code for Indoor Environmental Pollution Control of Civil Building Engineering" (2013 edition) and GB/T16147-1995 "Measurement Method for Radon Concentration in Air by Scintillation Bottle".
Instrument features:
◆ Small size, light weight, easy to carry.
◆ High sensitivity, low power consumption, dual-use for both AC and DC
◆ USB data transfer interface, Bluetooth printing.
Technical performance:
Sensitivity: ≥ 0.68cpm/[Bq • m-3]
Background count rate: ≤ 0.3cpm
The measurement range is as follows:
Environmental air radon: (3-100000) Bq/m ³
Soil radon: (300-300000) Bq/m ³
Radon release rate: (0.001~10.000) Bq/[m2S] (optional)
Radon in water: (0.003-100) Bq/L (optional)
Measurement repeatability error: ≤ 5% (radon chamber concentration 2000 Bq/m ³, environmental humidity 65%, temperature 25 ℃)
Long term stability: ≤ 10% (continuous operation for 8 hours)
Power supply: Lithium ion rechargeable battery/AC power, battery powered can work continuously for 30 hours
working environment
Temperature: -10 ℃ to+40 ℃
Humidity: Relative humidity ≤ 90% (+40 ℃)
Detector: ZnS (Ag) and photomultiplier tube combination system
Data storage: 2000 sets
Operation mode: single point detection or continuous monitoring
Display: LCD liquid crystal display
Gas extraction method: active pump suction
measure time
Air radon: 30 minutes
Soil radon: 11 minutes
Radon release rate: 30 min (optional)
Radon in water: 31 minutes (optional)
Host printing data: date, time, point number, and test results
Dimensions: Host: (330 × 210 × 170) mm
Weight: 5kg
Standard configuration:
(1) One instrument host.
(2) One power cord.
(3) One Bluetooth printer, print 2 rolls of paper.
(4) One set of soil sampler: including one steel drill rod, one sampler, two connecting hoses, and one drying tower.
(5) One copy of the instrument manual.
(6) One certificate. (Optional)
(7) Extract a set of accessories. (Optional)
(8) A set of radon water accessories. (Optional)