| Desktop pH 700 and Ion 700 testers Model: 35419 xx |
![]() |
| characteristic: Instrument Introduction Compact design, 40% smaller in size than other desktop testers Large screen display, convenient for reading Data recording function, allowing you to store and use up to 100 sets of data The Oakton desktop 700 series tester is specifically designed for use in crowded spaces such as laboratories. Compact and compact, saving valuable space, Dual row large screen display, capable of reading even from a distance. The pH700 tester can measure pH, oxidation-reduction potential, relative oxidation-reduction potential, and temperature. The Ion700 tester can also measure ion concentration, And display the measurement results directly in ppm units for ease of use. Automatic temperature compensation (ATC), even under fluctuating temperature conditions Ensure accurate readings. Automatic buffer recognition, compatible with USA (pH 1.68, 4.01, 7.00, 10.01, 12.45) and NIST Set 5 pH calibration values for pH buffer solutions (pH 1.68, 4.01, 6.86, 9.18, 12.45). The Ion700 tester also has an ion concentration calibration function, Five calibration values can be set.
Randomly equippedUniversal 100/240 VAC power adapter.
Technical Parameter
Slope control: 80-120%
Working temperature: 0~50 ℃
pH 700
35419-00(Testing only) 35419-03(Test meter with "fully functional" pH/ATC electrode) 35419-10(Test meter with double junction pH electrode, ATC probe, and detachable electrode holder) 35419-12(Test meter with detachable electrode holder) Range - pH: -2.00~16.00 Range - mV/Rel mV: ± 2000 Range - Temperature: 0.0~100.0 ℃
Resolution - pH: 0.01pH
Resolution - mV/Rel mV: 0.1 within ± 199.9mV range, 1mV beyond ± 199.9mV. Resolution - Temperature: 0.1 ℃ or ℉.
Accuracy - pH: ± 0.01pH
Accuracy - mV/Rel mV: ± 0.2 within the range of ± 199.9mV, ± 2 beyond ± 199.9mV. Accuracy - Temperature: ± 0.5 ℃
Calibration - pH: up to 5 calibration points, automatic buffer recognition.
Calibration - mV: intercept adjustment up to ± 150mV. Ion 700 35419-23(Test meter with "fully functional" pH/ATC electrode) 35419-20(Test meter with double junction pH electrode, ATC probe, and detachable electrode holder) 35419-22(Test meter with detachable electrode holder) Range - pH: -2.00~16.00 Scope - ISE: 0.01~2000ppm Range - mV/Rel mV: ± 2000 Range - Temperature: 0.0~100.0 ℃
Resolution - pH: 0.01pH
Resolution - ISE: 0.01~2000ppm Resolution - mV/Rel mV: 0.1 within ± 199.9mV range, 1mV beyond ± 199.9mV. Resolution - Temperature: 0.1 ℃ or ℉.
Accuracy - pH: ± 0.01pH
Accuracy - ISE: ± 5% of reading (unit price), ± 1% of reading (divalent). Accuracy - mV/Rel mV: ± 0.2 within the range of ± 199.9mV, ± 2 beyond ± 199.9mV. Accuracy - Temperature: ± 0.5 ℃
Calibration - pH: up to 5 calibration points, automatic buffer recognition.
Calibration - mV: intercept adjustment up to ± 150mV. |