English

Beijing Jiahua Zhongxin Technology Co., Ltd
Home>Products>Delger Multitest Medical Compressed Air Quality Tester
Firm Information
  • Transaction Level
    VIP member
  • Contact
  • Phone
    13671068345
  • Address
    Tiantong Xiyuan Third District, Dongxiaokou Town, Changping District, Beijing
Contact Now
Delger Multitest Medical Compressed Air Quality Tester
The MultiTest med. Int compressed air quality detector is used to detect the quality of breathing air supplied by compressors or compression cylinders
Product details

The MultiTest med. Int compressed air quality detector is used to detect the quality of breathing air supplied by compressors or compression cylinders. The detection system is reliable and meets the purification standard EN12021. All components are packaged in practical packaging and can be used immediately after disassembly. The use of Dr ä ger MultiTest med. Int for pollutant detection and measurement can reliably determine various potential pollutants in compressed air flow. The device is easy to use and fast to measure, and does not require a power supply. Instead, it can be easily connected to the low-pressure compressed air supply system to be monitored using a plug without any problems. The combination of medical gas detectors and Delger detection tubes can detect various substances in compressed air, such as nitrogen oxides, carbon dioxide, and oxygen. The Delger detection tube can be used independently for qualitative detection of water vapor, oil, CO2, SO2, H2S, NOX, CO, and other pollutants in medical gases, and can also be used for simultaneous detection. The detector can be connected to various types of quick plug interfaces, and the quick inspection function can provide accurate measurement results and display pollution levels in just five minutes.

Main application areas

Pharmaceutical industry, fire protection industry, spray painting industry, aerospace industry, electronic manufacturing industry, petrochemical industry, etc

Technical parameters:

Suitcase: 400mm in length, 300mm in width, and 91mm in height
Weight: 3 kilograms
Work pressure: 3-20 bar in accordance with DIN13260 standard
Interface: Quick plug interface G1/4“
Flow rate: 0.2 liters/minute and 4.0 liters/minute

MultiTest med. Int. ordering information:

Complete configuration 65 27 320

-1 oxygen pressure regulator (3-20 bar)

-1 flow regulator that can simultaneously install 7 detection tubes

-1 timer

-1 pipe opener

-1 user manual

-1 box of 10 Delger detection tubes, CO2 100/a-P

-1 box of 10 Delger detection tubes, carbon monoxide 5/a-P

-1 box of 10 Delger detection tubes, sulfur dioxide 0.5/a

-1 box of 10 Delger detection tubes, nitrogen oxide gas 0.5/a

-1 box of 10 Delger detection tubes, hydrogen sulfide 1/d

-1 box of 10 Delger detection tubes, oil 10/a-P

-1 box of 10 Delger detection tubes, water vapor 20/a-P

-1 handheld device

Excluding Delger detection tube 65 27 319

-1 oxygen pressure regulator (5 ± 0.5 bar)

-1 flow regulator that can simultaneously install 7 detection tubes

-1 timer

-1 pipe opener

-1 user manual

-1 suitcase

-1 oil tank adapter

(Additional testing tube according to requirements)


Supporting supplies

Name of detection tube

range

specifications

order No


carbon dioxide

100/a-P

10 pieces/box

6728521


carbon monoxide

5/a-P

10 pieces/box

6728511


sulfur dioxide

0.5/a-P

10 pieces/box

6728491


Nitrogen oxide gas

0.5/a

10 pieces/box

8103661


hydrogen sulfide

1/d

10 pieces/box

8101831


oil box

10/a-P

10 pieces/box

8103560


steam

20/a-P

10 pieces/box

8103061


Attention:

  • The oil detection box and detection tube must be used within the designated detection tube channel to ensure accurate measurement.

  • When using a detection tube, ensure that the direction of flow is correct.

  • Do not breathe during measurement.

  • After cleaning the measuring equipment, please ensure that each measuring unit is installed back to its corresponding connection position.

  • The instrument must be regularly inspected and maintained by professional technicians and recorded.

  • The maintenance work of the instrument needs to be carried out by trained service personnel.

  • Only use original Delger parts for repairs.

  • All components must be stored in a dry, cool, dust-free location, ensuring no direct sunlight or heating.

  • Physical display


Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!

Successful operation!