Functional features:
Ø1.approveFully automatic operation,Sample testing can be carried out continuously or with a fixed sample interval set, and the instrument can spontaneously and periodically measure samples on its own;
Ø2. It can be fully automated for cleaning, and can be manually cleaned at any time or after setting a fixed cleaning intervalSpontaneous periodic cleaning;
Ø3.Equipped with fully automatic calibration function,Calibration can be manually performed at any time or the instrument can calibrate itself after setting a fixed calibration cycle;
Ø4. Stable operation,No need for frequent maintenance,Maintenance interval>1 month, with each maintenance lasting approximately 30~60minute;
Ø5.Adequate memory,Storable20000 pieces of data,justNo loss in case of power failure, automatically overwrite the earliest data when fully stored(Customizable expansion of memory);
Ø6Adopting TFT true color touch screen, 7-inch large screen, 70000 colors, 800 * 480 resolution, the instrument's working status is clear at a glance;
Ø7. The instrument reserves a printer interface, which can be customized and equipped with industrial micro printers (optional);
Ø8. The instrument is equipped with a preprocessing system that can achieve functions such as self-cleaning, blowback, and precision filtration to avoid suspended particles blocking the pipeline.
Parameter indicators:
uMeasurement method:national standardGB11914-89, Potassium dichromate high-temperature digestion, colorimetric determination
uMeasurement range:0~200mg/L,0~1000mg/L,0~5000mg/LThree range options are available, maximum customizable 0~40000mg/L (automatic dilution)
uDetection limit:8mg/L
uresolving power:<1mg/L
uMeasurement accuracy: Standard solution<10%; Water sample<15%
uReproducibility:<5%
uAccessible operation time:≥ 720h/time
uRange drift:±5%F.S.
uSample interval: continuous,1 hour, 2 hours, 3 hours... 24 hours, triggered, specified time point
uSignal output:4-20mA (20mA corresponds to adjustable range); RS485 and RS232 are optional