Scope of Analysis |
Particle size: 0.2 µ m-1600 µ m |
Small hole tube diameter |
10 µ m -2000 µ m (nominal diameter) |
Dynamic range of small hole tube |
Standard: 1:30 (diameter); |
Resolution |
User Selectable |
Number of channels |
Digitized pulse data can select up to 400 analysis channels within any selected range. When necessary, the number and range of channels can be re divided. |
Electrolyte solution |
All recommended electrolyte solutions for small hole technology, both aqueous and non-aqueous, are suitable for Multisizer 4e. Electrolytes should be compatible with glass, fluoropolymers, fluoroelastomers, and stainless steel. |
Digital pulse processor |
Exclusive high-speed digital signal |
Pulse data |
Each single analysis has a time scale pulse of up to 525000 times |
Granularity distribution data |
Provide particle size distribution of diameter, volume, and area, including absolute quantity, quantity%, quantity/milliliter, volume, volume%, volume/milliliter, surface area, surface area%, and surface area/milliliter |
Pulse distribution data |
Provide pulse distribution with time, sequence, and width for pulse height diameter, pulse height volume, pulse height volt, pulse width, pulse area, average pulse height diameter, average pulse height volume, and average pulse width. Provide the quantity distribution of width. |
Linearity |
Diameter: ± 1%; Volume: ± 3% |
Small hole current range |
30 µ A -6000 µ A, step size: 0.2 µ A |
Small hole current accuracy |
± 0.4% of the set value |
Polarity error |
Less than 0.5% |
Time mode |
0.1-999 seconds, with selectable increments within 10 ms. Normally, time analysis ranges from 10 seconds to 90 seconds |
Total Count Mode |
50-50000 counts |
Single channel counting mode |
10-100000 counts |
Volume mode |
Available continuously from 50 microliters to 2000 microliters. |
Measurement system |
Mercury free, multi range metering pump |
Precision of positive displacement pump |
Better than 99.5% |
Compliance |
This software complies with the 21 CFR Part 11 standard |
Size and weight |
Net weight: 45 kg; |
Power frequency |
47 -63 Hz |
Power supply |
Less than 55 volt amperes (watts) |
Fuse type |
250 V, IEC (5 × 20 mm), Time Delay (TD), 2.0 A |
Environmental conditions |
To ensure safety, this instrument can only be used indoors; |
Working temperature |
5°C-40°C |
Relative humidity |
30% to 85%, no condensation |
Altitude |
Up to 2000 meters (6560 feet) |