The Agilent 7820A gas chromatograph is one of the stable gas chromatography system choices for small to medium-sized laboratories that use standard gas chromatography methods for routine analysis. seven thousand eight hundred and twenty
Can extend normal operating time, reduce maintenance requirements, lower complexity, and meet user expectations.
7820A gas chromatograph reference technical parameters
Size and weight:
Height x Width x Depth: 49cm x 56cm x 51cm
Average weight: 50 kg
Gas chromatography panel buttons and display: Chinese or English
Environmental requirements:
Working temperature: 15-30 ℃
Working humidity: 30~70%
Storage temperature: -40~70 ℃
Operating altitude: 3100 meters
Overall system performance:
Using 7820A, EPC control (no diversion), automatic sampler, Agilent workstation, analyze tridecane (2ng column loading), retention time reproducibility:<0.06%, peak area reproducibility:<2%
column oven
Size: 28.0 × 30.5 × 16.5 cm
Temperature range: 8 ℃~425 ℃ above room temperature
Temperature setting accuracy: 1 ℃
Heating rate: 75 ℃/min
50-70 ℃, heating rate: 75 ℃/minute
70-115 ℃, heating rate: 45 ℃/minute
115-175 ℃, heating rate: 40 ℃/minute
175~300 ℃, heating rate: 30 ℃/minute
300-425 ℃, heating rate: 20 ℃/minute
Running time: 999.99 minutes
Program heating: 5th order