100kNElectro hydraulic servo dynamic and static universal testing machine
1. Equipment Overview
1.1 Name, Model, and Specifications
Name: Electro hydraulic servo dynamic and static universal testing machine
model: LF5105
Specification: Maximum load of 100kN
1.2 Function Overview
The LF5105 electro-hydraulic servo dynamic static universal testing system from Lishi Scientific Instrument Company is used for testing applications of general materials. The LF series floor standing loading frame is an ideal choice for conducting tests ranging from material mechanical properties to small component testing. It is suitable for testing applications on various materials, such as composite materials, steel, aluminum alloys, superalloys, and more.
This scheme includes a 100kN loading framework, which is equipped with advanced digital controllers and professional application software in the industry. It is also equipped with fatigue test specific hydraulic fixtures, crack propagation test fixtures, extensometers and other test accessories.
The design and manufacturing of this system equipment should comply with ISO/ASTM international standards, and all components, instruments, and sensors should be measured using the International System of Units (SI) standard.
1.3 Relevant standards
ØASTM D3479 Pull fatigue test method,
ØASTM D7615 Open Hole Tensile/Compression Fatigue Test Method
ØGB/T 3075 Metallic Axial Fatigue Test Method
ØHB5287 Metallic Materials Axial Loading Fatigue Test Method
ØGB/T 15248-2008 Metallic Materials Axial Equal Amplitude Low Cycle Fatigue Test Method
ØGB/T 6398-2000 Test Method for Fatigue Crack Growth Rate of Metallic Materials
ØGB/T 4161-2007 Metallic materials - Plane strain fracture toughness KIC test method
ØGB/T21143Unified test method for quasi-static fracture toughness of metallic materials
ØASTM 1290Test method for measuring crack tip opening displacement (CTOD) and crack toughness
ØASTM-E399 Metallic Materials Linear Elastic Plane Strain Fracture Toughness KIC Test Method
ØJJG 556-2011Axial force fatigue testing machine
2. TECHNICAL INDEX
Specification and model |
LF5105 |
Maximum load (kN)) |
100 |
Accuracy level of testing machine |
Static indication accuracy: ± 0.5%; Dynamic loading accuracy: ± 2%; |
Relative error of test force indication |
Static ± 0.5%Within |
Maximum amplitude of actuator |
±75 mm |
Displacement measurement accuracy |
±0.5%FS |
Displacement measurement resolution |
0.001mm |
Relative error of deformation indication |
±0.5%Within |
sampling frequency |
10kHZ |
Frequency range |
0.01-50Hz |
Main experimental waveforms |
Sine wave, square wave, triangular wave, oblique wave, random waveform, and external input waveform; Can achieve sine waveforms with multiple combinations of different frequencies or amplitudes; Users can customize the random waveform of parameters; User imported custom waveforms |
Experimental control method |
Load, displacement, and deformation control methods |
Dynamic test line speed |
The maximum testing speed for the experiment is 50mm/s |
Maximum counting range |
109 |
counting error |
≤±1order |
coaxality with load |
≤5% |
Load sensors have the ability to resist overload |
300%, fatigue level. |
Effective distance between columns (mm)) |
540mm |
Maximum experimental space (mm)) |
1100mm(Excluding hydraulic fixtures) |
Specification of constant pressure servo pump station |
internet traffic40L/min,21Mpa, Motor Power30kW |
Weight (kg) |
1200 |