VIP member
WK-2000 microcoulometric comprehensive titrator
Brief description: The WK-2000 microcoulometric comprehensive titrator applies the principle of microcoulometric titration and computer-controlled dat
Product details
Microcoulometric comprehensive titratorsummary:
WK-2000 modelMicrocoulometric comprehensive titratorApplying the principle of microcoulometric titration, computer-controlled data acquisition and processing. The sample is injected into the lysis tube by an automatic sampler for reaction, and the tested element is converted into titratable ions, which are carried into the titration cell by a carrier gas for titration. The titrant in the electrolyte is consumed, and an equal amount of titrant is generated and consumed. The amount of electricity consumed during the electrolysis process is measured, and the content of the tested element in the sample is calculated according to Faraday's law.
According to different configurations, the content of sulfur or chlorine in the sample can be measured separately. The WK-2000 model complies with standards such as SH/T0253, SH/T0254, and SH/T0222.
Technical parameters:
1. Measurement range: Sulfur 0.2mg/L to 10000mg/L
Chlorine 0.5mg/L~10000mg/L
2. Sample types: solid, liquid, gas (with corresponding injector)
3. Temperature control range and accuracy: Room temperature to 1000 ℃, ± 0.5% ± 5 ℃
WK-2000 repeatability error:
X<1.0mg/L ± 0.1mg/L
1.0mg/L ≤ X ≤ 10mg/L LCv ≤ 10%
X> 10mg/LCv ≤ 5%
Online inquiry