Instrument features:
★The display window adopts5.7Designed with an inch industrial color LCD screen, it displays more comprehensive information and has a more reasonable interface operation;
★Having proficiency in both Chinese and English2Set up an operating system to meet different user needs;
★Abandoning fragile and low-end productsPVCLeather covered buttons, made of plastic molds, have a good tactile feel and are durable;
★Adopting advanced technology10/100MAdaptive Ethernet communication interface, built-inIPProtocol stack, facilitating long-distance data transmission for enterprises through internal LAN and the Internet; Facilitating laboratory setup, simplifying laboratory configuration and data management;
★Internal Design3An independent connection thread that can connect to local processing, unit supervisors (such as chief engineers, technical plant managers, etc.), and higher-level supervisory departments (such as environmental protection bureaus, technical supervision bureaus, etc.), facilitating real-time monitoring of instrument operation and analysis of data results by unit supervisors and higher-level supervisory units;
★Equipped withNETChrom®Workstation, capable of supporting multiple chromatography instruments(253Simultaneously working, achieving data processing and counter control, reaching a leading level in the industry;
★ NETChrom®Built in workstationModbus/TCPServer, which can easily access the analysis resultsDCS(Distributed Control System;
★Adopting a modular structural design, the design is clear and easy to replace and upgrade, protecting the effectiveness of investment. It can meet the analysis of complex samples and can be equipped with various high-performance detectors, such asFID、TCD、ECD、FPDandNPDWait;
★Completely abandons traditional pointer pressure gauges and can be loadedEPCTechnology for air path control, with automation level and overall performance close to international top brands;
★Implemented self-protection against gas path faults, automatic ignition and shutdown, and automatic opening of the gas path, achieving one key start;
★Design a timed self starting program that can easily complete online analysis of gas and liquid samples (with injection components required);
★The system is designed with an automatic sampler interface and comes with multiple built-in driver programs, allowing for the installation of an automatic sampler at any time;
Main technical indicators:
★Interface display:5.7Inch industrial color LCD screen
★Temperature controlled area:8road
★Temperature control range: above room temperature4℃~450℃Incremental:1℃Accuracy:±0.01℃
★Program heating order:8rank
★Cheng Sheng rate:0.1~39℃/min(Ordinary type);0.1~80℃/min(High speed type)
★External events:6Road; Auxiliary control output2road
★Sample injector types: packed column injection, capillary injection, six way valve gas injection, automatic headspace injection optional
★Number of detectors:3(maximum);FID、TCD、ECD、FPDandNPDoptional
★Gas path control: mechanical valve control modeEPCMethod selection
★ EPC、EFCWorking mode:2Seed; Constant current mode, constant voltage mode
★ EPC、EFCWorking gas:5Seed; Nitrogen, hydrogen, air, helium, argon
★ EPC、EFCCheng Sheng:4rank
★ PC、EFCControl range: Pressure:0~0.6MPa; Traffic0~100mL/minperhaps0~500mL/minair
★Pressure sensor:
nAccuracy: Full scale<±2%
nReproducibility:<±0.05 KPa
nTemperature coefficient:<±0.01 KPa/°C
★Flow sensor:
nAccuracy: Full scale<±5%
nReproducibility:<±0.5%(full scale)
nRange:0~500mL/min
★Start injection: manual or automatic selection
★Communication interface: Ethernet:IEEE802.3
★Power Supply:220V±10%,50Hz;2500W(Maximum)
★Volume:572×552×465(High)mm
★Weight:50kg(About)
★Technical specifications of detector
nHydrogen flame ionization detector(FID)
uDetection limit:Mt≤3×10-12g/s(Hexadecane)-Isooctane solution);
uNoise:≤5×10-14A
uDrift:≤1×10-13A/30min
uLinear range:≥106
nthermal conductivity detector(TCD):
uSensitivity:S≥3500mV•ml/mg(Conventional)5000mV•ml/mg(High sensitivity)(benzene-Toluene solution)(Zoom in)2、4、8Multiple Choice)
uNoise:≤10μV
uBaseline drift:≤30μV/30min
uLinear range:≥104
nECd(ECD):
uDetection limit:≤1×10-14g/s
uNoise:≤0.03mV
uBaseline drift:≤0.2mV/30min
uLinear range:≥103
uRadioactive Source:63Ni
nflame photometric detector(FPD):
uDetection limit:(S)≤5×10-11g/s,(P)≤1×10-12g/s;
uNoise:≤0.03mV
uBaseline drift:≤0.2mV/30min
uLinear range:≥103(S), one hundred and two(P)