Intelligent features
-
Adjustable backlit display screen for easy viewing
-
Flexible speed adjustment
-
Optional 6-liter pure water tank, no need to directly connect the system to the water pipeline
Inlet water quality monitoring
-
High quality incoming water quality ensures a continuous supply of ultrapure water and maximizes the lifespan of the filter element
Clever design
-
The recirculation pump can prevent the growth of bacteria during standby and continuously monitor the limit value. When an error occurs during operation, an error message will be displayed
-
All required pressure reducing valves, ultrapure water filter cartridges, and 0.2 µ m sterile filters for each system are fully equipped
Excellent filtration system
-
The ultrafiltration device can automatically flush to maximize the retention of endotoxins and nucleases, with a service life of up to two years.
-
Verified 0.2 µ m final filter with pleated membrane for up to five times disinfection effectiveness
-
RNase:0.003ng/ml; Dnase:0.4pg/μl
MicroPure Basic analytical type |
MicroPure UV Organic analytical type |
MicroPure UF Biochemical analysis type |
MicroPure UV/UF Life Science Type |
|
application |
AA,IC,ICP, Buffer solution, etc |
Inorganic measurement analysis, organic measurement analysis, HPLC, ICP-MS, IC, TOC analysis |
Molecular Biology, Microbiology, PCR,IVF, Monoclonal antibodies, etc |
Molecular Biology, PCR,DNA, Monoclonal antibodies, cell culture, etc |
Electrical resistivity @ M Ω - cm | 18.2 | |||
Conductivity μ s/cm | 0.055 | |||
TOC ppb | 5 - 10 | 1 - 5 | 5 - 10 | 1 - 5 |
Bacterial CFU/ML | < 1 | |||
Particle size 0.22 μ m/mL | < 1 | |||
Endotoxin EU/mL | n/a | n/a | < 0.001 | < 0.001 |
Flow rate L/min | ≤ 1 | |||
Power (KW) | 0.06 | |||
External dimensions Width x Depth x Height (mm) |
305 × 300 × 455 | |||
External dimensions (with water tank) Width x Depth x Height (mm) |
305 × 400 × 455 | |||
Item number | ||||
Add a 6-liter water tank |
Water intake demand | |||
Source water | Reverse osmosis water, deionized water, distilled water | ||
Inlet conductivity (μ s/cm) | < 5 | ||
TOC (ppb) | ≤ 50 | ||
Bacteria (CFU/mL) | < 100 | ||
Turbidity NTU | < 1.0 | ||
Temperature (℃) | 2 - 35 | ||
Pressure (bar) | 0.1 - 6 |
*Please refer to the user manual for more detailed information on the required inlet water quality
Select accessories |
|
||||
Product Name | Description | Item number | |||
Ion exchange column DI 1500 |
If RO water inlet is used to extend the ultra purification column afterwards service life |
02.1500 |
|||
printer | Compliant with GLP standards, RS-232 interface printer | 09.2207 | |||
MicroPure IQOQ certification |
IQ OQ certification | IOQDOCE50133913 | |||
MicroPure+30L Pure Water Tank IQOQ Certification | IQ OQ certification | IOQDOCE50133914 | |||
Wall mounted components | Optional accessories for hanging MicroPure on the wall | 09.2212 | |||
Disinfection column |
To disinfect the MicroPure pure water system, a disinfection kit is required Use together |
09.1102 |
|||
Common consumables |
|
||||
Ultra purification column | MicroPure uses ultra purification ion exchange column | 09.1006 |
|
||
Terminal 0.2 μ m Microfiltration | Used for filtering microorganisms and particulate matter in ultrapure water terminals | 09.1003 |
|
||
Ultrafiltration filter cartridge | Used for removing endotoxins, nucleases, microorganisms, particles, etc |
|
|||
UV lamp | 185/254 nm dual wavelength ultraviolet lamp tube | 09.1002 |
|
||
Disinfection bag | Disinfectant | 09.2202 |
|