Karl Fischer volumetric moisture analyzer MKV-710S

The main features of the Karl Fischer volumetric moisture analyzer MKV-710S are:

1. Adopting a large 8.4-inch color LCD touch screen for operation and control.
2. The touch screen and the host adopt a separate design and can be displayed simultaneously.
3. The measuring unit can be placed inside a fume hood, which is safer and reduces the risk.
4. User permission setting function to prevent incorrect settings and facilitate control.
5. The new burette unit can reduce dead volume and reagent usage.
6. The intelligent burette unit stores the information of the titrant in the chip.
7. The measurement results can be stored on a USB flash drive and generate PDF experimental reports.
8. Connect the ADP-611 furnace to automatically determine the appropriate heating temperature.

Technical parameters of Karl Fischer volumetric moisture analyzer MKV-710S:

Name and Model: Karl Fischer Moisture Analyzer MKV-710S.
Instrument composition: MCU-710S+MKV-710+solvent exchange unit (optional).
Measurement method: Karl Fischer titration method.
Measurement range: 1) Moisture content: 0.1-500mg H2O, 2) Moisture concentration: 1ppm~100% H2O.
Burette accuracy: burette volume: 10mL, accuracy: ± 0.015mL, repeatability: ± 0.005mL.
Endpoint determination: Determine the endpoint when the duration of the polarized potential is within the specified range, with endpoint time ranging from 1 to 99 seconds.
Titration form: forward titration/reverse titration (additional purchase of a second set of burettes is required).
Solvent demand: 30mL to 100mL (S-type titration cup).
Measurement method: 120 sets of measurement methods.
Input settings: Touch screen input.
Display: 8.4-inch color LCD screen, available in eight languages including Chinese, English, Japanese, Korean, Russian, Spanish, German, and French, with one channel display.
Calculation: concentration calculation, statistical calculation, automatic input of blank value and titration degree.
Data storage: 500 sets of sample results.
GLP certification: Registered operator/user group management, titrant management, performance testing, time management.
External output: RS-232C (printer/balance/data software/Karl Fischer oven), USB (USB flash drive/thermal printer/A4 printer/keyboard/barcode machine/foot switch/USB hub), SS-BUS(APB)。
Expansion function: Second set of burette driver unit, moisture evaporator ADP-611.
Usage environment: Temperature: 5-35 ° C, relative humidity: Below 85% RH.
Power Supply: AC100-240V ± 10%, 50Hz/60Hz.
Power consumption: Host: about 30 watts, printer: about 7 watts.
Size: Touchscreen: 225 (W) × 190 (D) × 42 (H) mm, titration unit: 141 (W) × 296 (D) × 367 (H) mm, mixer: 107 (W) × 206 (D) × 322 (H) mm.
Weight: Touch screen: about 1.5 kilograms, titration unit: about 4.0 kilograms, mixer: about 2.0 kilograms.