German WTW Turb 750T turbidity analyzer (SF free shipping) Product Introduction:
Applications: Water, wastewater, wine, beer, food, livestock/aquaculture, pharmaceuticals, industrial
There are two types of instruments to comply with the regulations on drinking water supervision:
Turb 750IR according to DIN EN ISO 7027-1
According to US EPA 180.1 Turb 750T
German WTW Turb 750T turbidity analyzer (SF free shipping)Product Features
1、 The Turb 750 series is calibrated using validated AMCO Clear standards and utilizes the best measurement window on the vial. The instrument comes with a calibration kit for 3-point calibration. The following menu guidance calibration function is available:
1) Default 3-point calibration
2) Flexible calibration using 2-5 user-defined calibration points
3) Quick calibration: 1-point calibration
4) Setting of calibration interval
5) Storage of Calibration Protocol
2、 Advantages of AMCO Clear Standard
The AMCO Clear standard is made of polymer microspheres and has the following accuracy and precision:
1) Long term stability, no drift (different from formalin)
2) The accuracy between batches is 1%
3) The optical components have been optimized separately
4) Can be traced back to Fumalazine
5) Environmentally friendly and non-toxic
6) Easy to use
A successful calibration using AMCO Clear standard is necessary to obtain accurate and reproducible results on Turb 750.
3、 Obtain reliable results using AQA
1) Easy to clean keyboard
2) AQA supports: interval, tolerance, protocol
3) Direct access to basic functions
4) AQA flag in the dataset
5) Storage with sample ID
6) Calibration interval and calibration protocol
7) Data filter for data output
8) GLP compliant data management
4、 Using Turb Data for Data Management
The measured values are stored as a dataset along with the associated calibration protocol, sample ID, and AQA information. Data output can be done through a printer or by pressing the print button on Hyperterminal. Alternatively, using the provided PC software Turb Data can make it easier to complete:
1) GLP compliant data transmission
2) Instrument identification through serial number
3) Username input
4) Output in CSV format
5) Export data to Excel
6) Import data into LIMS
Technical parameters:
Instrument model | Turb 750IR | Turb 750T |
Measurement standards | DIN EN ISO 7027-1 | US EPA 180.1 |
light source | Infrared light emitting diode | White tungsten filament lamp |
Measurement mode | Scattering photometry method(90° Scattered light) | |
display | Backlit graphic display,160*104pixel | |
keyboard | Easy to clean foil keyboard with alphanumeric input options | |
range | 0-1100 FNU/NTU | 0-1100 NTU |
unit | FNU/NTU | NTU |
resolution |
exist0.00-9.99 FNU/NTUWithin the range0.01 FNU/NTU exist10-99.99 FNU/NTUWithin the range0.1 FNU/NTU exist100-1100 FNU/NTUWithin the range1 FNU/NTU |
|
accuracy | 0.01 FNU/NTU or ± reading2% |
0.01NTU or ± reading2% exist500-1100NTUWithin the range, the reading is ±3% |
reproducibility | <Reading0.5% | <Reading1% |
reading modes | Adopting only reproducibility and reliability checks(IRPC)Program, fast response time, andIRCPSupport rapid settling samples for measurement. | |
response time | 4 seconds | 7 seconds |
calibration-option | default3Point standard calibration and use2-5Flexible and rapid calibration of user-defined calibration points | |
Calibration protocol and interval setting | correct/correct | |
AQS support | Calibration protocolAQASigns, calibration marks | |
Data Storage | 2500 datasets with calibration protocolAQAsign | |
Sample identification | Enter letters and numbers through the keyboard | |
Software Update | pass throughUSB | |
interface | RS 232、USBThe printer passes throughPCperhapsRS 232 | |
Instrument storage conditions | -25-+65℃(13-149°F)Work | |
Working temperature range |
+5-+55℃(41-131°F) When connecting the power plug, it is+5-+40℃(41-104°F) |
|
conform toGLPofPCsoftware | Turb Data | |
Size (height)*wide*Deep) | make an appointment290*190*80mm(11.42*7.48*3.15Inches) | |
weight | 1.1kg | |
Calibration standard | used for3Calibration kit for point standard calibration: Long term stable polymerAMCO ClearStandard products,0.02-10.0-1000 FNU/NTU | |
Colorimetric bottle, sample size | The diameter is28mmA small bottle with a small capacity15mlBorosilicate glass, phenolic resin bottle cap, PTFE coated rubber seal. Silicone oil is not required to cover scratches in the measurement program. | |
Sample conditions | Sample temperature<70℃(158°F) | |
power supply | Wide range power supply commonly used for plugs in Europe, the United States, the United Kingdom, and Australia | |
certificate | CE | |
configuration | Laboratory turbidity meterTurb 750 IR/TFour sections1.5VAABattery, wide range power supplyUSB-AtoUSB-B miniCable, six empty spaces28mmSmall bottle, bottle cap label for directional marking, calibration kitTurb 430/750 IRperhapsTCloth, quick guide, concise operation manual, with extended manualCD-ROMWithTurb Data PCSoftwareCD-ROMInspection Protocol |