Beijing Haiguang Instrument Co., Ltd
Home>Products>GGX-910 Zeeman flame atomic absorption spectrophotometer
Firm Information
  • Transaction Level
    VIP member
  • Contact
  • Phone
    13331195206
  • Address
    Building M3, No.1 Jiuxianqiao East Road, Chaoyang District, Beijing
Contact Now
GGX-910 Zeeman flame atomic absorption spectrophotometer
Atomic absorption spectrophotometer is one of our company's traditional products, with more than 20 models put into the market. The instrument has sta
Product details
Performance characteristics

1. Zeeman background correction, full band background subtraction, constant magnetic field, transverse Zeeman technique;
2. Quasi dual beam, no preheating required;
3. Eight light automatic lighthouse, simultaneously preheating eight elemental lights;
4. A patent (ZL 2018207884083) design for a multi position rotating lighthouse lighting system, which overcomes the rotation and entanglement of light cables;
5. Wavelength range 190-900nm; nm;
6. Czerny turner type optical path design, 1800 engraved lines/mm plane diffraction grating;
7. Focal length 270mm, short optical path, strong energy;
8. The spectral bandwidth is automatically adjusted to 0.2, 0.4, 1.0, and 2.0;
9. Automatic peak setting and automatic spectral bandwidth;
10. Chinese Win software, with powerful functions, allows for the selection of any calibration curve and any blank for calibration.


characteristic

1. Automatic gas circuit, using a mass flow meter, with continuously adjustable gas flow rate;

2. Titanium alloy combustion head, all plastic atomization chamber, strong corrosion resistance;

3. High sensitivity glass atomizer;

4. Automatic water seal detection, ignition is prohibited when the water seal is short of water;

5. Equipped with waste liquid level detection to prevent waste liquid from overflowing;

6. Equipped with air and acetylene pressure monitoring, ignition is prohibited if the pressure is insufficient;

7. Equipped with ignition safety inspection function, complete safety interlock protection device and anti backfire device;

8. Equipped with flame monitoring, the gas will automatically shut off in case of accidental shutdown;

9. Can be equipped with a hydride device to measure arsenic, mercury, etc.


Patent technology:

Zeeman background correction optical path and system ZL 2015100677596

A multi position rotating lighthouse lighting system ZL 2018207884083


Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!

Successful operation!