The F-series miniature high-pressure fluorine pump adopts a small-sized design that is conducive to system integration. The pump head is made of 316L stainless steel embedded with PCTFE design, and is matched with high-pressure resistant PTFE pipe to achieve a pressure resistance of 5MPa, effectively combining small volume, corrosion resistance, and high pressure resistance functions.F-series miniature high-pressure fluorine pumpIt also has online flow and pressure change curves, client flow correction, volumetric and weight based delivery functions,Modbus RTUandASCIIagreement, RS232/RS485/usb,PLCAnalog input:0—5V , 4—20 mAWeighing and metering system for reactor and chemical reaction device pump inlet weight reduction; High flow accuracy, low pressure pulse, good repeatability, excellent quality, widely used in microreactors, chemical reaction devices, catalytic evaluation devices for chemical feeding, reactant addition, catalyst addition, as well as supporting devices for supercritical fluid transportation, foaming extrusion processes, new energy research, catalytic refining kettle reactions, etc.
Patent Name: A High Pressure Resistant Fluoroplastic Infusion Pump Patent Number: 0
F0505A high-pressure fluorine pumpTechnical features:
1. Pressure protection setting: The maximum and minimum pressure limits can be set;
2. Low pressure pulse, using cam curve compensation and flow pulse electronic suppression technology;
3. Switching flow rate units: delivering by volume in ml/min and by mass in g/min;
4. Flow calibration: Provide a user parameter area for calibrating the flow rate of the pump to meet the requirements of different system operating conditions;
5. Flow control program: provides two types of running programs, constant current running and gradient running;
6. Pressure curve display: Each device displays the current pressure curve separately, allowing for a visual assessment of the device's operational status;
7. Flow curve display: Each device displays the current running flow curve separately;
8. Piston rod cleaning system, extending the service life of the pump;
9. Supports RS-232, RS-485, RS-422, USB communication; Modbus RTU and ASCII protocols;
10. One to many control, free selection of single pump operation and multiple gradient operation, expandable 1-to-32 operation;
11. Running data saving: Start or stop data recording, and save the pressure curve and flow curve to an Excel format file.
12. Setting method saving: Save the running parameter settings, such as serial port parameters, pressure protection values, flow control programs, etc.
13. It can achieve closed-loop control of the fluid inlet balance weight reduction system (customized according to customer balance specifications).
F0505A high-pressure fluorine pumpTechnical parameters:
Serial number | Product model | F0505A |
1 | Pump head material | Stainless steel embedded PCTFE pump head, threaded on stainless steel |
2 | Infusion method | Dual plunger parallel mode, floating plunger design |
3 | Flow range | 0.01-50.00ml/min |
4 | increment | 0.01ml/min |
5 | Traffic accuracy | ± |
6 | 0.5% | Traffic repeatability |
7 | ≤ | 0.1% |
8 | Pressure range | ≤ |
9 | 5Mpa | pressure fluctuation |
10 | ≤ | 0.05Mpa |
11 | Sealing ring structure | Open sealing ring |
12 | Flow path material | PCTFE、PTFE、 Ruby, ceramics, Hastelloy |
13 | Export pipeline | φ3mm、φ3.175、φ4mm |
14 | Traffic correction point | 0.01, 0.1, 1, 10, 20, 40, 60, 80, 100ml/min |
15 | Flow correction function | User end |
16 | Multi parameter flow correction function within the range of 0.01 to 100.00ml | Communication function |
17 | RS232, USB (standard); RS485/422 (optional); | Modbus RTU and ASCII protocol |