E+H FMI52 Cable type Capacitive Level GaugeAdvantages:
The Liquicap FMI52 probe can be used for continuous measurement of liquid level, with reliable measurement, especially suitable for small storage tanks, media with adhesive formation, and extreme high temperature situations. The measured value is independent of the DK value (dielectric constant). The E+H FMI52 cable capacitive level gauge is used in conjunction with the Fieldgate FXA320 (remote measurement value access using internet technology), and Liquicap provides an ideal inventory management solution.
advantage
The E+H FMI52 cable type capacitive level gauge does not require calibration for liquids with conductivity greater than 100 μ S/cm. The probe length is pre-set at the factory at 0%/100%, and debugging is fast and simple.
Text display, menu guided on-site settings (optional).
We have obtained various certifications and licenses, with a wide range of applications.
Compliant with IEC 61508 standard, suitable for safety systems that require SIL2 level safety.
The liquid receiving part is made of corrosion-resistant materials and FDA certified materials.
The E+H FMI52 cable type capacitive level gauge has dual overvoltage protection (gas discharge device+protective diode) to prevent tank discharge.
Electronic components provide built-in adhesion compensation
application area
The Liquicap FMI52 has a sturdy structure (self sealing cone) and a probe pressure range from vacuum to 100 bar. The temperature range of sealing and isolation materials used for storage tanks can reach -80~+200 ℃. When the conductivity of a liquid is greater than 100 μ S/cm, the measured value is independent of DK (dielectric constant), so there is no need to recalibrate to measure different liquids.
E+H FMI52 Cable type Capacitive Level GaugeTechnical parameters:
measuring principle |
capacitance |
Features/Applications |
Fully insulated cable probe, suitable for standard and extreme process conditions (temperature, pressure) |
Interface measurement |
Interfaces liquid / liquid also with emulsion layers |
Power/Communication |
12-36V DC HART,PFM |
measurement accuracy |
repeatablity 0,1% |
Linear error in measuring conductive liquids |
<0,25% |
ambient temperature |
|
process temperature |
|
Process absolute pressure/maximum overpressure limit |
Vacuum can reach 100 bar (vacuum can reach 1450 psi) |
Contact medium components |
Insulation materials: PTFE, PFA, 316L |
Process connection |
'G 3/4, G 1, G1 1/2 /NPT 1/2,NPT 3/4', NPT 1', NPT1 1/2; DN25.../ANSI 1 '.../JIS... flange |
Sanitary process connection |
ISO02852 three clamp head, milk connector, flush installation seal |
sensor length |
Total length: 6m (20ft); Shielding section length: maximum 2m (7ft) |
Maximum measurement distance |
0.42 m ... 10.0 m (1.38 ft ... 33 ft) |
output |
4...20mA HART PFM |
Certificates and certifications |
ATEX,WHG,SIL,EHEDG |
optional |
Separated shell; Sealed airtight probe |
characteristic |
Interface measurement; Shielding segment length; Factory calibration; Short measurement response time |
Application limitations |
Pay attention to the corresponding pressure and temperature values, applicable to DK for non-conductive media |