The industrial online residual chlorine monitoring instrument is a water quality online monitoring and control instrument with a microprocessor. This
Product details



※ Electrode installation methods: flow-through, pipeline, and submerged
The CL2160 series industrial online residual chlorine meter is a water quality online monitoring and control instrument with a microprocessor. This instrument is widely used in various industries such as drinking water treatment plants, drinking water distribution networks, swimming pools, water quality treatment engineering, sewage treatment, and water disinfection (with chlorine dioxide generators) to continuously monitor and control the residual chlorine and temperature values in aqueous solutions.
Instrument features
● LCD large screen LCD display
● Chinese intelligent menu operation
● Multiple automatic calibration functions
● Signal output mode optional
● Manual and automatic temperature compensation
● Manual pH compensation function
● Two sets of relay control switches
● High limit, low limit, hysteresis control
RS485//RS232 communication connection is optional
● Display residual chlorine, pH, and temperature values on the same interface
● Password protection is provided to prevent non staff personnel from operating incorrectly
Main technical parameters
(1) Measurement range (can be manually set):
Residual chlorine: 0-2.00mg/L (ppm);
0~5.00mg/L;
0~10.00mg/L;
0~20.00mg/L;
Temperature: 0-50.0 ℃;
(2) Resolution:
Residual chlorine: 0.001mg/L; Temperature: 0.1 ℃;
(3) Basic error:
Residual chlorine: ± 1% F · S;
Temperature: ± 0.3 ℃;
(4) Response time: 90% less than 90 seconds;
(5) PH compensation range:
Manual pH compensation: 5-9pH;
(6) Isolation current output:
0-10mA (load resistance<1.5K Ω);
4-20mA (load resistance<750 Ω);
(7) Two sets of relay control contacts:
3A 240VAC, 6A 28VDC or 120VAC;
(8) Instrument power supply:
220VAC ± 10%, 50 ± 1Hz, power ≤ 3W;
(9) Instrument dimensions: 96 × 96 × 130mm;
(10) Instrument installation method: disk mounted (embedded);
Installation hole size: 91 × 91mm;
(11) Optional features:
RS485//RS232 communication interface connected to computer;
(12) Instrument weight: 0.6kg;
(13) Instrument working environment:
Environmental temperature:- 10~60℃;
Relative humidity: not exceeding 90%;
There is no strong magnetic field interference around except for the Earth's magnetic field.