The instrument adopts microcomputer technology for timing control heating method, and can simultaneously heat the all glass reflux device of 250ml conical (straight large grinding mouth) bottles with 6/8/10/12 grinding ports; To achieve the goals of energy conservation, reducing power load, and improving efficiency.
The instrument uses glass burr reflux tubes instead of spherical reflux tubes, and replaces tap water cooling with air and water cooling technologies, which can save water, reduce storage space, standardize the instrument, and improve the reliability of instrument use; The heating bottom plate adopts a fashionable microcrystalline optical panel design concept, which not only improves the accuracy of sample digestion, but also enhances the beauty of the product.
The chemical solution preparation, operation, and COD calculation of the instrument fully comply with the national standard HJ 828-2017; Note: For COD water samples with a concentration of ≤ 50mg/L, the sampling volume can be appropriately increased when the sample concentration is low; For COD water samples with a concentration greater than 50mg/L, the accuracy can be improved by diluting the titrant and oxidant; For COD water samples with a concentration greater than 700mg/L, dilution of the water sample concentration is required to complete the result determination; For water samples with high chloride ion content, the method in Appendix A of the national standard HJ 828-2017 can be used for determination.
Dual fan convection cooling, independently controllable, with significant cooling effect
Large screen LCD Chinese display, adjustable time and power, users can modify, set, and save on their own
Adopting microcrystalline heating panel and infrared heating method, the heating speed is fast and uniform, and the service life is longer
After digestion is completed, the instrument automatically sounds a warning and stops heating work
Water saving, energy-saving, low thermal radiation, and small space occupation
Easy to operate, can effectively reduce costs and improve work efficiency
Measurement range: 4-700mg/L, 700-1000mg/L (requires diluted water sample)
Measurement time: ≤ 150 minutes (can be set according to user needs)
Measurement error: Potassium hydrogen phthalate standard solution (500mg/L), relative standard deviation not exceeding 5.0%, industrial organic wastewater (500mg/L), relative standard deviation not exceeding 8.0%
Environmental temperature:- 5~40℃
Display mode: LCD display
Dissolved samples: 8 per batch
Dissolving time: can be set according to user needs, 1-999 minutes
Fan settings: automatic/manual
Heating power: 1200W (AC 220V ± 22V, 50HZ)
Burr reflux tube (8 sets), power cord, instruction manual, certificate of conformity, etc