GSL-1700X-II is a CE certified dual temperature zone high-temperature tube furnace, with each heating zone length of 12 '', controlled by a 30 segment intelligent temperature regulator. When the temperature reaches 1700 degrees, the heating element uses silicon molybdenum rods, and when the temperature reaches 1400 degrees, the heating element uses silicon carbon rods. By adjusting the temperature control program of the two temperature zones, a temperature gradient can be formed in the temperature field inside the furnace tube. This series of tube furnaces can use CVD or PVD methods to grow nanomaterials and produce various thin films.
Technical Parameter | |
---|---|
shell structure |
Double layered shell structure with air cooling system, ensuring that the surface temperature of the shell is less than 60 degrees Celsius The furnace insulation material adopts high-purity alumina polycrystalline fibers to reduce heat loss The surface of the inner furnace is coated with a high-temperature alumina coating imported from the United States, which can improve the heating efficiency of the equipment and also extend the service life of the instrument |
power |
9KW |
voltage |
AC 380V , 50 Hz |
Heating zone length |
Two temperature zones are heated separately ● Temperature Zone 1: Operating Temperature 800-1400 º C (Silicon Carbon Rod Heating) Heating zone length: 12" (300 mm) Length of constant temperature zone: 4" (100mm) within +/-1ºC Heating rate: ≤ 10ºC/min ● Temperature Zone 2: Operating Temperature 800-1700 º C (Silicon Molybdenum Rod Heating) Heating zone length: 12" (300 mm) Length of constant temperature zone: 4" (100mm) within +/-1ºC Heating rate: ≤ 5ºC/min Note: (Constant temperature zone) Due to the influence of measurement methods and external environment, the measurement results may have deviations. This parameter is for reference only and should not be used as a technical indicator for this equipment. You need accurate constant temperature zone data to purchase the product. Please contact our sales personnel |
Furnace tube size and flange |
99.8% purity alumina ceramic tube ● (Standard) (60mm) O.D. x (54mm) I.D x 1000 mm Length Including 60mm stainless steel vacuum sealed flange, needle valve, and pressure gauge |
Expansion |
If you want to increase the pumping rate to obtain a higher vacuum, it is recommended that you replace the standard hose connector with the KF25 adapter adapter (Click to view) If you need to use it under inert or reducing gas conditions, it is recommended that you replace the standard hose connector with a VCR double sleeve connector (click to view) ● Optional: Our company produces hinged flanges and imported anti-corrosion vacuum gauges
|
temperature control |
PID intelligent temperature controller, capable of setting up 30 temperature rise and fall programs Each temperature zone can be independently set with overheat and trip protection ●Temperature control instrument operation video |
thermocouple |
A B-type thermocouple and an S-type thermocouple |
Control method |
PID automatic control |
precision |
● +/- 1ºC |
heating element |
● Temperature zone 1: 8 pcs U type SiC rods (1500C grade ) ● Temperature Zone 2: 8 pcs U MoSi2 rods (1800C grade ) |
External dimensions |
780 L x 450 W x 720 H mm |
net weight |
150 kg |
Warranty period |
One year warranty period, excluding consumables such as heating elements, sealing rings, furnace tubes, and other consumables |
Certificate |
● CE certification All electrical components (>24V) are certified by UL/MET/CSA If the customer incurs certification fees, our company guarantees that a single device will pass the German TUV certification or CAS certification |
national patent |
Patent Name: A Wide Range Operating Temperature Openable High Temperature Tube Furnace Patent Number: ZL-2011-2-0389859. 8 Respect innovation, disdain plagiarism, and hold infringement accountable |
Application precautions |
The air pressure inside the furnace tube cannot exceed 0.02MPa Due to the high internal pressure of the gas cylinder, a pressure reducing valve must be installed on the cylinder when introducing gas into the furnace tube. It is recommended to purchase a pressure reducing valve from our company, which has a range of 0.01MPa-0.1MPa and can be used more accurately and safely When the temperature of the furnace body is higher than 1000 ℃, the furnace tube cannot be in a vacuum state. The pressure inside the furnace tube should be equivalent to atmospheric pressure and maintained at atmospheric pressure The gas flow rate entering the furnace tube should be less than 200SCCM to avoid the impact of cold large airflow on the heated quartz tube The long-term use temperature of quartz tube is less than 1100 ℃ It is not recommended to close the exhaust and intake valves at the flange end of the furnace tube for sample heating experiments. If it is necessary to close the gas valve to heat the sample, it is necessary to constantly pay attention to the reading on the pressure gauge. If the pressure reading is greater than 0.02 MPa, the vent valve must be opened immediately to prevent accidents (such as furnace tube rupture, flange flying out, etc.) |